Laboratory education
Research Peptide Testing Glossary
Use this reference when a label, certificate of analysis or safety document uses terms that look similar but answer different questions.
Lot or batch
An identifier for a defined quantity of material. A testing lab may use a different sample ID. The sender should be able to show how that sample maps to the lot on the label.
COA and specification
A certificate of analysis records tests and their results for a sample. A specification states the requirements a result is compared against. They are not interchangeable.
Identity, purity and assay
Identity addresses what the material is. Purity describes what the stated method measures in relation to other material or signals; an HPLC area percentage is not automatically a mass fraction. Assay typically addresses content or amount under a stated method.
Endotoxin and sterility
An endotoxin test addresses bacterial endotoxins. It does not establish sterility. A result reported as below a threshold should not be read as zero.
Lyophilization
Another word for freeze-drying. It describes a processing step, not the synthesis location, identity, purity or testing status of a reagent.